Basic Edition


Overview


A 96-element curve-linear array collects more than 69,000 detection points in a single 360° rotation, resolving 150 µm in transverse planes at depths of 20–40 mm. Excitation covers 532 nm and 650–1320 nm, switchable to any wavelength in range every 50 ms.

Oxyhaemoglobin, deoxyhaemoglobin and melanin are visible without a contrast agent, so anatomy and vasculature are recorded alongside every molecular scan.

The chassis, holders, scan times and software are shared with the Premium edition. Basic has no fluorescence channel.

Specifications


Photoacoustic (PA) imaging channel
Type3D high-resolution deep tissue molecular, physiological and anatomical imagingSubcutaneous and skin imaging
Spatial resolution150 µm × 150 µm; 150 µm × 420 µmTransverse planes; sagittal and coronal planes
Molecular imaging sensitivity100 nM ICGIn vitro tests, CNR = 2
PA excitation range532 nm and 650–1320 nm
Detection points per scan> 69,000Single scan, 360° azimuthal rotation
Detector configurationCurve-linear arrayCylindrical focusing
Detector central frequency6 MHz ± 10%T/R measurements, optimised sensitivity in receive mode
Number of array elements96Wide-angle 3D imaging transducers
PA signal digitizerLEGION ADC12-bit, 256 parallel channels, up to 400 Hz frame rate, 40 MHz sampling, programmable amplifier 46–91 dB
Laser excitation unit
Tunable wavelength range532 nm and 650–1320 nm
Pulse repetition frequency20 Hz
Pulse energy> 130 mJ @ 700 nm; > 10 mJ @ 532 nmBefore fiber bundle transmission
Energy meterReal-time automatic pulse energy measurement
Fast wavelength switchingAny wavelength between 650–1320 nm every 50 ms
Chassis typeMobileOn wheels, positioned next to the image acquisition unit
Dimensions (L × W × H)68 cm × 44 cm × 89 cm
Power requirements208 or 240 VAC, single phase 50/60 Hz, < 1.5 kVA
Image acquisition unit
Standard scan time36 s360° azimuthal rotation, 720 data frames
Scan typesContinuous azimuthal rotation or reverse scans (≤ 360°)Time-limited to 10 min
Excitation sequenceSingle wavelength; linear or custom wavelength sweepSpectral unmixing presets for molecular, physiological and anatomical imaging
Max single-scan 3D image50 mm × 50 mm × 30 mm
Whole-body imagingStack of 3D volumes, manual axial positioning10 mm steps, 40 mm positioning range, 50 × 50 × 70 mm total imaging range
In vivo imaging subjectsMice, rats (< 200 g)Fur should be shaved or depilated from the studied section before imaging
Max weight of test subject0.5 kg
Coupling mediumDI waterSubject submerged under anesthesia, degassing enabled
Environment temperature control20–40 ± 0.5 °CControlled heating and circulation of the coupling liquid
Test subject monitoringVisual monitoring with a camera
Laser safetyLight-tight imaging chamber, laser interlocksNo eye protection required
Chassis typeBenchtop
Dimensions (L × W × H)70 cm × 35 cm × 70 cm55 cm × 35 cm footprint
Power requirements208–240 V 4 A or 120 V 8 A, 50/60 Hz
Control station
Form factorDesktop mid-tower or Mini ITX case
ConfigurationHigh-performance Nvidia GPU, SSD, Windows 10/111440p or higher monitor, keyboard, mouse
Imaging softwareTriTom Imaging Suite; 3D SlicerAcquisition, reconstruction and molecular imaging; visualisation and analysis of 3D volumes
Data formatsScan data: raw, mat. 3D image: vtk, mat
Excitation fiberoptic bundle
Transmission> 70%
Excitation spot, axial size30 mm
Length2 m

Accessories included


Supplied with every system
Gas anesthesia systemMice and small rats; includes animal induction chamber
Mouse restrainerB-type for the abdominal region and legs; H-type for the thoracic region, head and neck
Microcuvette holderScans up to ten 50 µL cuvettes of liquid sample, quick setup
MicrocuvettesCylindrical PTFE, 0.8 mm ID, 50 µm wall thickness, for 50 µL samples
Coupling liquid containersFill and drain the image acquisition unit